Design of measurement system of lens center thickness based on double-side optical confocal technology

  • Received Date: 2015-08-04
    Accepted Date: 2015-10-13
  • For the existing measurement system of lens center thickness based on optical confocal technology, refraction and reflection of incident light affected the measuring accuracy. In order to solve this problems, based on confocal technology, the modified measurement system of lens center thickness was designed with the method of double-side optical confocal measurement. After theoretical analysis, error analysis and experimental verification, the results show that the measurement range of the system reaches 30mm and the measurement accuracy is 2m. The system can get the high-speed untouched measurement of center thickness and satisfy the demands of actual measurement.
  • 加载中
  • [1]

    LIANG H F, CAI P F. Non-contact measurement method for thickness of optical element[J]. Optical Technique, 2014, 40(6):535-538(in Chinese).
    [2]

    XIONG Zh Y, ZHAO B. Design of a new laser thickness measurement instrument[J].Laser Technology, 2011, 35(5):614-617(in Chinese).
    [3]

    YU D Y, TAN H Y. Engineering optics[M]. 3rd th.Beijing:Mechanical Engineering Press,2011:33-37(in Chinese).
    [4]

    GAO M,HUANG D J,LIU J.The measurement instru-ment of center thickness of lens[J].Journal of Xi'an Technological University, 2003, 23(3):234-236(in Chinese).
    [5]

    QIAO Y, ZHANG N, LIU T, et al. Optical system design for the measurement of thickness of lens based on confocal technology[J]. Optical Technique, 2010, 36(6):857-859(in Chinese).
    [6]

    MATTHIAS K, JOCHEN S. Noncontact measurement of central lens thickness[J]. Glass Science and Technology, 2005,78(5):245-247.
    [7]

    LU Y, XU X P, SHI N, et al. Research on measurement method and apparatus for lens center thickness[J]. Journal of Changchun University of Science and Technology (Natural Science Edition), 2013, 36(3/4):29-31(in Chinese).
    [8]

    HUANG S,DENG L Y,YANG H, et al. Homogenization design of laser diode based on ZEMAX[J].Laser Technology, 2014, 38(4):523-526(in Chinese).
    [9]

    ZHU W B,ZHONG J,MO R Y,et al.Design of spectral confocal design of spectral confocal chromatic displacement sensor objective[J]. Opto-Electronic Engineering,2010,37(8):62-66(in Chinese).
    [10]

    QIAO Y, ZHANG N, XU X P. Design of lens thickness measurement system based on confocal technology[J]. Chinese Journal of Scientific Instrument, 2011, 32(7):1635-1641(in Chinese).
    [11]

    YANG M S, XING P F, GAO D Z, et al. Measurement of thickness of foils based on conjugated confocal micro-displacement sensors[J]. Atomic Energy Science and Technology, 2010, 44(s1):577-578(in Chinese).
  • 加载中
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Article views(5203) PDF downloads(478) Cited by()

Proportional views

Design of measurement system of lens center thickness based on double-side optical confocal technology

  • 1. School of Mechanical Engineering, Jiangsu University, Zhenjiang 212013, China

Abstract: For the existing measurement system of lens center thickness based on optical confocal technology, refraction and reflection of incident light affected the measuring accuracy. In order to solve this problems, based on confocal technology, the modified measurement system of lens center thickness was designed with the method of double-side optical confocal measurement. After theoretical analysis, error analysis and experimental verification, the results show that the measurement range of the system reaches 30mm and the measurement accuracy is 2m. The system can get the high-speed untouched measurement of center thickness and satisfy the demands of actual measurement.